Module 6
Low Current Measurement
6-15
Approx. 50 sec
< +/- 3 fA
< +/- 3 fA
> +/- 3 fA
ZERO CHECK
Effect of connecting a triaxial cable
If you change anything in your fixturing or cabling, wait at least 1 minute before making a
measurement at low currents.
Here we see the effect of connecting a cable to the open SMU port of the B1500A. The discontinuity
lasted for 50 seconds of a 2 minute sweep from 0 to 1 volt in 10 mV steps.