Agilent B2900 Users Guide, Edition 2 4- 19
Front Panel Reference
Display and Assist Keys
NOTE If the number of data exceeds 5000
If the number of measurement data exceeds 5000, the Graph view and the Roll view
plot the following data. where, n is integer, 1 to 5000.
Number of data is 5001 to 10000: [2*(n-1)+1]th data
Number of data is 10001 to 25000: [5*(n-1)+1]th data
Number of data is 25001 to 50000: [10*(n-1)+1]th data
Number of data is 50001 to 100000: [20*(n-1)+1]th data
Roll View
Available on B2911A and B2912A. Displays the time domain graph for plotting the
channel 1 and/or 2 measurement data. See
Table 4-3 for the Y-axis data types. The
Roll view can plot up to 5000 data.
Display area 1. Displays status ON or OFF, on B2911A
Displays status Ch 1, Ch 2, or OFF, on B2912A
Two lines can be displayed on the graph. [1] indicates the graph setup for line 1.
[2] indicates the graph setup for line 2.
2. Y-axis data type I, V, R, or P.