6- 12 Agilent B2900 Users Guide, Edition 2
Function Details
Over Voltage/Current Protection
Over Voltage/Current Protection
The over voltage/current protection is effective for preventing damage to the test
device from overcurrent or overvoltage. If this function is enabled, the
source/measure unit (SMU) sets the output to 0 V and sets the output switch to off
automatically and immediately when it reaches the compliance status.
To set the over voltage/current protection, see “Output Connection dialog box” on
page 4-25.