Graphical view of limits and specifications
As with other measurements the actual limits depend on many factors, namely,
class, type, system and power level. Figure 16 gives example limits for E-GSM900
normal BTS at high power.
Practical measurements
Spectrum due to switching measurements are less difficult and less demanding
than spectrum due to modulation and wideband noise measurements. In
practice, equipment that can perform the latter can easily manage the former.
When to use the measurement
Spectrum due to switching measurements are usually performed alongside
spectrum due to modulation and wideband noise measurements.
f
c
+ 400 kHz
f
c
+ 600 kHz
f
c
+ 1200 kHz
f
c
+ 1800 kHz
Freq (MHz)
f
c
= 942.5
Power
(dBc)
57 dBc
67 dBc
74 dBc
74 dBc
Referenced to power
in 300 kHz RBW
Trace shown in
30 kHz RBW
Example: E-GSM900, normal BTS, middle channel, high output power
Channels and slots:
B, M, T. Single carrier
All slots on and alternate
Hopping:
Off and On (B, M, T)
RBW:
30 kHz
VBW:
100 kHz
Detection:
Peak hold
Zero span
Filter:
5-pole sync tuned
Notes:
For low O/P power levels
further conditions apply (limits
are less demanding).
P
max
P
max
30 dB
P
idle
Figure 16. Spectrum due to switching, BTS, limits