Practical measurements
As mentioned, modern test equipment performs the necessary signal processing
automatically, making these measurements straightforward and fast. It is also
useful to view phase error versus time—especially in R&D and when fault
finding. For example, a phase and frequency error test might fail the prescribed
limits at only one point in the burst, for example, at the beginning. This could
indicate a problem with the transmitter power ramp or some undesirable
interaction between the modulator and power amplifier.
Constellation diagrams can also be used to observe some aspects of modulation
accuracy and can reveal certain fault mechanisms such as I/Q amplitude
imbalance or quadrature imbalance.
When to use the measurement
Phase and frequency error measurements can capture a large spread of
fault types and prove that any I/Q calibration process has been successfully
performed. These measurements are typically used at every stage in the BTS
lifecycle. Modern test equipment can make these measurements very rapidly
and with good accuracy (typically the test equipment should be 10x more
accurate than the specification limit so measurement results can be attributed
to the device under test and not the test system).
Figure 5. Quad display of the Agilent E4406A VSA-series transmitter tester with and without frequency error
Figure 6. Constellation diagram on the Agilent E4406A VSA-series transmitter tester