Chapter 4: Installing Oscilloscope Measurement Modules
16533/34A Oscilloscope Module (single or multi-card modules)
16533/34A Oscilloscope Module (single or multi-card
modules)
The Agilent Technologies 16533A/34A oscilloscope module functions as either
a master card or expander card. It is compatible with all versions of software
in your 16700A/B or 16702A/B mainframe. The circuitry in the module
requires an operational accuracy calibration to optimize measurement
accuracy.
A multicard module should contain either all 16533A or 16534A cards.
NOTE: Each of the individual cards of a multicard 16533A or a multicard 16534A
module must first be calibrated as a single card. After reconfiguring into a
multicard module, the channel skew calibration needs to be done.
Step 1
Prepare a single 16533/34A card for calibration
CAUTION: The effects of ELECTROSTATIC DISCHARGE can damage components. Use
grounded wrist straps and mats when you are performing any kind of service
on this module.
1 Power down your 16700A/B or 16702A/B mainframe.
2 Remove all modules and filler panels from the mainframe and set the
PROTECTED/UNPROTECTED switch to UNPROTECTED.
16700A
Power
Busy
16700B or 16702B