78
Chapter 4: Installing Oscilloscope Measurement Modules
16533/34A Oscilloscope Module (single or multi-card modules)
Step 3
Calibrate a single card for external trigger skew
1 Connect a 9-inch 50-ohm BNC cable to one side of the BNC tee
adapter. On the other side of the BNC tee adapter connect a 9-inch 50-
ohm BNC and a BNC(f)/SMB(m) adapter.
2 Select the Procedure field and then select Ext Trig Skew.
Channel 1
ECL
AC/DC Cal
BNC/SMB
Adapter
External
Trigger In