1. For Option 521, maximum reverse power is ½ watt when Option 1E1 step attenuator is set at or above 5 dB. When Option 1E1 step attenuator = 0 dB, or for units
without Option 1E1, maximum reverse power is 2 watts above 250 MHz, 1/2 watt below 250 MHz.
Adjustable RF output limit (with Option 521)
14
0246 8101214161820
16
18
20
22
24
26
28
30
32
Frequency [GHz]
Power [dBm]
15dBm
20dBm
25dBm
30dBm
Off
Leveling modes Internal leveling, external detector leveling, millimeter source module, ALC off
External detector leveling
Range –0.2 mV to –0.5 V (nom) (–36 dBm to +4 dBm using Agilent 33330D/E detector)
Bandwidth Selectable 0.1 to 100 kHz (nom) (Note: not intended for pulsed operation)
Maximum reverse power 1/2 Watt, 0 V
DC
1
Adjustable RF output limit
Function Protects external devices by limiting maximum RF output. Operates in all leveling modes
(internal, external, source module)
Range User-adjustable from +15 dBm to maximum output power
Accuracy
+15 to +25 dBm ±1 dB (typ)
> +25 dBm ±1.5 dB (typ)
Resolution 1 dB
Response time 30 μsec (measured)
Adjustment Can be locked to prevent accidental change
Option 521
10 to < 500 MHz < 6:1 (typ) without Option 1E1, or step attenuator set to 0 dB
< 1.6:1 (typ) with Option 1E1 step attenuator set ≥ 5 dB
0.5 to 20 GHz < 1.8:1 (typ)
RF output limit (measured)
11