14
16
18
20
22
24
26
28
30
32
34
00.5 11.5 2
Option 521 maximum output power
Harmonic filters - on
Harmonic filters - off
Frequency [GHz]
Power [dBm]
Option 532/540 with Options 1EU, 1E1 (measured)
16
18
20
22
24
26
28
Frequency [GHz]
Power [dBm]
0510 15 20 25 30 35 40
Option 550/567 with Options 1EU, 1E1
0
5
10
15
20
25
30
Frequency [GHz]
Power [dBm]
010203040506
0
Step attenuator (Option 1E1)
1
Options 520, 521, 532, and 540 0 dB and 5 dB to 115 dB in 10 dB steps
With Optimize S/N on
2
0 dB to 115 dB in 5 dB steps
Options 550 and 567 0 dB to 90 dB in 10 dB steps
Attenuator hold range
Minimum From –20 dBm to maximum specified output power with step attenuator in 0 dB position. Can be offset using
Option 1E1 attenuator.
1. The step attenuator provides coarse power attenuation to achieve low power levels. Fine power level adjustment is provided by the ALC (automatic level control)
within the attenuator hold range.
2. With attenuator in auto mode. Optimize S/N mode provides improved signal/noise performance and is included with Option 521 and Option 1EU models. Specs in
the following sections (such as level accuracy, spectral purity, modulation, etc) are only tested with Optimize S/N mode turned off.
Maximum output power (measured)
20 GHz models maximum output power
14
16
18
20
22
24
26
28
30
32
34
Frequency [GHz]
Power [dBm]
02468101214161820
Opt 520 - STD, 1E1
Opt 520 - 1EU, 1E1
Opt 521 - 1E1
Opt 521
Option 520-1EU, 1E1 maximum output power
15
17
19
21
23
27
00.5 11.5 2
25
Filters on
Filters off
Frequency [GHz]
Power [dBm]
8