Agilent Technologies E8257C PSG Manual de usuario Pagina 3

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 35
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 2
Specifications
1. For Option 521, performance is degraded below 500 MHz. Refer to specifications for more detail.
2. In ramp sweep mode (Option 007), resolution is limited with narrow spans and slow sweep speeds. Refer to ramp sweep specifications for more information.
3. Time from GPIB trigger to frequency within 0.1 ppm of final frequency above 250 MHz or within 100 Hz below 250 MHz. CW switching speed to within 0.05% of final
frequency is ≥ 5 ms (nom).
4. Add 12 ms (typical) when switching from greater than 3.2 GHz to less than 3.2 GHz.
5. With Option 1EH low band harmonic filters off. With the 1EH filters turned on, add 4 ms.
6. N is a factor used to help define certain specifications within the document.
7. Not verified by Agilent N7800A TME Calibration and Adjustment Software. Daily aging rate may be verified as a supplementary chargeable service, on request.
Frequency bands Frequency range N
6
1 250 kHz to 250 MHz 1/8
2 > 250 to 500 MHz 1/16
3 > 500 MHz to 1 GHz 1/8
4 > 1 to 2 GHz 1/4
5 > 2 to 3.2 GHz 1/2
6 > 3.2 to 10 GHz 1
7 > 10 to 20 GHz 2
8 > 20 to 40 GHz 4
9 > 40 GHz 8
Accuracy ± [(time since last adjustment x aging rate) + temperature effects + line voltage effects + calibration
accuracy]
Internal timebase reference oscillator
Aging rate
7
< ±3 x 10
–8
/year or
< ±2.5 x 10
–10
/day after 30 days
Initial achievable calibration accuracy < ±4 x 10
–8
Temperature effects (typ) < ±4.5 x 10
–9
from 0 to 55 °C
Line voltage effects (typ) < ±2 x 10
–10
for ±10% change
Frequency
Range Specified range Tunable range
Option 520 250 kHz to 20 GHz 100 kHz to 20 GHz
Option 521
1
10 MHz to 20 GHz 10 MHz to 20 GHz
Option 532 250 kHz to 31.8 GHz 100 kHz to 31.8 GHz
Option 540 250 kHz to 40 GHz 100 kHz to 40 GHz
Option 550 250 kHz to 50 GHz 100 kHz to 50 GHz
Option 567 250 kHz to 67 GHz 100 kHz to 70 GHz
Resolution
CW 0.001 Hz
All sweep modes
2
0.01 Hz
CW switching speed
3, 4, 5
Standard Opt UNX Opt UNY
< 11 ms (typ) <11 ms (typ) < 26 ms (typ)
< 7 ms (nom) < 7 ms (nom) < 22 ms (nom)
Phase offset Adjustable in nominal 0.1 ° increments
3
Vista de pagina 2
1 2 3 4 5 6 7 8 ... 34 35