Spectral purity
Harmonics
1
(dBc at +10 dBm or maximum specified output power, whichever is lower)
Frequency Options 520, 532, 540, 550, 567 Option 521
< 1 MHz -25 dBc (typ)
1 to < 10 MHz -25 dBc
10 to < 60 MHz -28 dBc -25 dBc
10 to < 60 MHz with Option 1EH filters on -45 dBc
2
-35 dBc
2, 3
0.06 to 2 GHz -30 dBc -25 dBc
0.06 to 2 GHz with Option 1EH filters on -55 dBc
2
-35 dBc
2, 3
> 2 to 20 GHz -55 dBc -35 dBc
> 20 to 67 GHz -50 dBc (typ)
1. Specifications are typical for harmonics beyond specified frequency range (beyond 50 GHz for Option 567). Specifications are with Option 1EH Low-pass Filters
below 2 GHz off and Option UNX or UNY low phase noise mode off unless noted.
2. Below 250 MHz in ramp sweep mode (Option 007), Option 1EH filters are always off. Refer to harmonic specification with filters off.
3. Option 521 includes low-pass filters below 2 GHz as standard.
4. -45 dBc below 60 MHz.
10 to 250 MHz, Option UNX or UNY low phase noise mode
With Option 1EH filters off -8 dBc (typ) -8 dBc (typ)
With Option 1EH filters on -55 dBc
4
-35 dBc
12