1. To within 0.1 dB of final amplitude within one attenuator range. Does not apply to Option 521 below 500 MHz.
2. To within 0.5 dB of final amplitude within one attenuator range. Also applies to Option 521 below 500 MHz with ALC on. Add up to 50 ms when using power
search.
3. Specifications apply in CW and list/step sweep modes over the 15 to 35 ºC temperature range with the ALC on. Degradation outside this temperature range, for
power levels > -10 dBm is typically < 0.3 dB (except < 0.5 dB from 2 to 3.2 GHz and with Option 521 below 500 MHz). In ramp sweep mode (with Option 007),
specifications are typical. For instruments with Type-N connectors (Option 1ED), specifications are degraded typically 0.2 dB above 18 GHz.
4. When Option UNX or UNY low phase noise mode is on, specifications below 250 MHz apply only when Option 1EH low-pass filters below 2 GHz are on. With
Option 1EH low-pass filters below 2 GHz off, accuracy is typically ±2 dB.
5. For Option 550 and 567, degrade level accuracy by 0.2 dB from 1.7 to 2 GHz when step attenuator is set to 0 dB or when Option 1E1 is not present.
6. Nominal above +16 dBm from 10 MHz to 60 MHz.
7. With Option 521, specifications below 500 MHz are typical, and apply for a 50 Ω load with VSWR less than 1.4:1.
8. Typical above +26 dBm.
9. Typical below -15 dBm.
10. Specifications apply in CW and list/step sweep modes over the 15 to 35 ºC temperature range, with the ALC on and attenuator hold off (normal operating mode).
Degradation outside this temperature range, with attenuator hold on and ALC power levels > -10 dBm, is typically < .3 dB (except < 0.5 dB from 2 to 3.2 GHz
and with Option 521 below 500 MHz). In ramp sweep mode (with Option 007), specifications are typical. For instruments with type-N connectors (Option 1ED),
specifications apply to 18 GHz only. From 18 to 20 GHz, typical level accuracy degrades by 0.2 dB. Specifications do not apply above the maximum specified
power.
11. With Option 521, specifications below 500 MHz apply with step attenuator set to 5 dB or higher (requiring Attenuator Hold ON above 8 dBm). With step attenua-
tor set to 0 dB, refer to level accuracy specifications without Option 1E1.
Level accuracy
3
(dB) > 20 dBm 20 to > 16 dBm 16 to > 10 dBm 10 to > 0 dBm 0 to –10 dBm < –10 to –20 dBm
Options 520, 532, 540, 550, 567
250 kHz to 2 GHz
4, 5
±0.8 ±0.8
6
±0.6 ±0.6 ±0.6 ±1.2
> 2 to 20 GHz ±1.0 ±0.8 ±0.8 ±0.8 ±0.8 ±1.2
> 20 to 40 GHz ±1.0 ±1.0 ±0.9 ±0.9 ±1.3
> 40 to 50 GHz ±1.3 ±0.9 ±1.2
> 50 to 67 GHz ±1.5 ±1.0 ±1.2 (typ)
Option 521
10 to < 500 MHz
4, 7
±1.9 (typ) ±1.2 (typ) ±1.2 (typ) ±1.1 (typ) ±1.2 (typ) ±1.2 (typ)
0.5 to 20 GHz ±1.0
8
±0.8 ±0.8 ±0.8 ±0.9 ±1.1
9
Amplitude switching speed
ALC on < 6 ms (typ)
1
ALC off <10 ms (typ) (not including power search)
2
Level accuracy with step attenuator (Option 1E1)
10
(dB)
26 to > 20 dBm 20 to > 16 dBm 16 to > 10 dBm 10 to > 0 dBm 0 to –10 dBm < –10 to –70 dBm < –70 to –90 dBm
Options 520, 532, 540, 550, 567
250 kHz to 2 GHz
4, 5
±1.0 ±0.8
6
±0.6 ±0.6 ±0.6 ±0.7 ±0.8
> 2 to 20 GHz ±1.0 ±0.8 ±0.8 ±0.8 ±0.8 ±0.9 ±1.0
> 20 to 40 GHz --- ±1.0 ±1.0 ±0.9 ±0.9 ±1.0 ±2.0
> 40 to 50 GHz --- --- --- ±1.3 ±0.9 ±1.5 ±2.5
> 50 to 67 GHz --- --- --- ±1.5 ±1.0 ±1.5 (typ) ±2.5 (typ)
Option 521
10 to < 500 MHz
4, 11
±1.3 ±0.8 ±0.8 ±0.7 ±1.0 ±1.0
0.5 to 20 GHz ±1.0
8
±0.8 ±0.8 ±0.8 ±0.8 ±1.1 ±1.1
9