Agilent Technologies J-BERT N4903B Manual de servicio Pagina 103

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 406
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 102
[SOURce[1]]:PATTern:EADDition:SOURce[?]
IAgilentN490xPGErrorAddition.Mode (not IVI-compliant)
[SOURce[1]]:PATTern:EADDition:SOURce EXTernal | FIXed
[SOURce[1]]:PATTern:EADDition:SOURce?
EXT | FIX
The command controls the source of injected errors:
EXTernal (and :EADDition[:STATe] is ON)
Each pulse at the Error Add port causes an error to be added to the data stream.
FIXed (and :EADDition[:STATe] is ON)
Repetitive errors are internally added to the data stream. The rate of error
addition is controlled by the :EADDition:RATE command.
The query returns the current error addition mode.
[SOURce[1]]:PATTern:EADDition:SELect[?]
[SOURce[1]]:PATTern:EADDition:SELect DATa | AUXiliary | BOTH
[SOURce[1]]:PATTern:EADDition:SELect?
The command is used to select the channel for error insertion. It could be either
Data or Auxiliary or both.
The query returns the currently selected error insertion channel.
N O T E
Each channel is using its own error insertion logic, that is triggered individually
when inserting an error. As an result of this, the errors inserted on Data Out and
Aux Data Out are normally not affecting the same bits on both channels.
[SOURce[1]]:PATTern:FORMat[:DATA][?]
IAgilentN490xPGPatternfile.SetData (IVI-compliant)
[SOURce[1]]:PATTern:FORMat[:DATA] <PACKed>, <NR1>
[SOURce[1]]:PATTern:FORMat[:DATA]?
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 103
IVI-COM Equivalent
Syntax
Return Range
Description
Syntax
Description
IVI-COM Equivalent
Syntax
Vista de pagina 102
1 2 ... 98 99 100 101 102 103 104 105 106 107 108 ... 405 406

Comentarios a estos manuales

Sin comentarios