the number of errors detected (NR3)
Passed/failed flag (NR1); 0 = failed, 1 = passed
MEAS:JTOL:CHAR(*):FETC:DATA:CAPability?
IAgilentN490xJCharacterizationMeasurement.GetCapbability()
MEAS:JTOLerance:CHAR(*):FETC:DATA:CAPability? DUT | INST
This query returns a comma-separated list that contains one value for each
measured frequency.
The values depend on the query parameter:
INSTrument: The maximum jitter amplitude the Serial BERT could generate.
DUT: The maximum jitter amplitude the device under test could stand without
exceeding the target BER.
The unit is UI.
5 SCPI Command Reference
364 Agilent J-BERT N4903B High-Performance Serial BERT
IVI-COM Equivalent
Syntax
Description