OUTPut[1]:HOLD[?]
IAgilentBertPGPulse.DelayHoldMode
OUTPut[1]:HOLD <WIDTh | DCYCle>
OUTPut[1]:HOLD?
Determines whether Width or Duty Cycle shall be kept, if a repetitive pulse
waveform (as in RZ or R1 signal modes) is enabled and the generator's signal
frequency is changed.
If this command is used to switch from "Hold Width" to "Hold Duty Cycle", the
current pulse width is converted to a percentage of the present signal period.
If this command is used to switch from "Hold Duty Cycle" to "Hold Width", the
current duty cycle is converted from percent to seconds, according to the present
signal period.
The query returns the current setting as a string.
OUTPut[1]:POLarity[?]
IAgilentN490xPGOutput.Polarity (IVI-compliant)
OUTPut[1]:POLarity NORMal | INVerted
OUTPut[1]:POLarity?
NORM | INV
The command sets the polarity of the pattern generator's Data Out port. The query
returns the current polarity of the pattern generator's Data Out port.
OUTPut[1][:STATe][?]
IAgilentN490xPGOutput.Enabled (IVI-compliant)
OUTPut[1][:STATe] 0 | 1 | OFF | ON
OUTPut[1][:STATe]?
5 SCPI Command Reference
138 Agilent J-BERT N4903B High-Performance Serial BERT
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Return Range
Description
IVI-COM Equivalent
Syntax
Vista de pagina 137