OUTPut2:CENTer
IAgilentN490xPGGlobal.OutputsDisconnect (not IVI-compliant)
OUTPut2:CENTer DISConnect | CONNect
The DISConnect command sets the voltage at the pattern generator's Clock Out,
Data Out , Aux Data Out and Trigger/Ref Clock Out ports to 0 V, the CONNect
command re-enables the output (to the normal data pattern).
This command is identical with “OUTPut[1]:CENTral[?] ” on page 135.
OUTPut2:COUPling[?]
IAgilentN490xPGClockVoltageTerminationEnabled
OUTPut2:COUPling AC | DC
OUTPut2:COUPling?
The command enables or disables the source of the termination voltage:
DC: Enables the termination voltage
AC: Disables the termination voltage
The query returns the current state.
OUTPut2[:STATe][?]
IAgilentN490xPGClock2Enabled
OUTPut2[:STATe] 0 | 1 | OFF | ON
OUTPut2[:STATe]?
This command is kept for compatibility reasons. Due to the lack of relays, it is not
possible to disable the Serial BERT's output. This command has no effect. The
query will always return "1" = ON.
To force the Data Out and Clock Out ports to 0 V, use the command “OUTPut
[1]:CENTral[?] ” on page 135.
5 SCPI Command Reference
172 Agilent J-BERT N4903B High-Performance Serial BERT