[P]FETCh[:SENSe[1]]:ERATio:OASZero[:TOTal]?
This is a contraction of the phrase One received AS Zero. The query returns the
ratio of erred ones (a true data one received a data zero) to number of bits.
[P]FETCh[:SENSe[1]]:ERATio:ZASone[:TOTal]?
IAgilentN490xEDErrorRatio.ReadZAO (IVI-compliant)
[P]FETCh[:SENSe[1]]:ERATio:ZASone[:TOTal]?
This is a contraction of the phrase Zero received AS One. The query returns the
ratio of erred zeros (a true data zero received a data one) to number of bits.
[P]FETCh[:SENSe[1]]:SFECount Subnode
This subnode has the following SCPI structure:
This subnode has the following commands:
Table 77
Name Description under
[:ALL][:FULL][:TOTal]? “[P]FETCh[:SENSe[1]]:SFECount[:ALL]
[:FULL][:TOTal]? ” on page 300
[:ALL][:FULL]:DELTa? “[P]FETCh[:SENSe[1]]:SFECount[:ALL]
[:FULL]:DELTa? ” on page 301
[P]FETCh[:SENSe[1]]:SFECount[:ALL][:FULL][:TOTal]?
[P]FETCh[:SENSe[1]]:SFECount[:ALL][:FULL][:TOTal]?
5 SCPI Command Reference
300 Agilent J-BERT N4903B High-Performance Serial BERT
Syntax
Description
IVI-COM Equivalent
Syntax
Description
Syntax