Service 8
Agilent 3000 Series Oscilloscopes User’s Guide 137
Procedure
1 Preset and calibrate the power meter according to the instructions
found in the power meter manual.
2 Set up the Power Meter to display measurements in units of Watts.
3 On the oscilloscope, press Save/Recall
.
4 Select the Storage item in the Save/Recall menu until “Setups” appears.
Oscilloscope
Power Meter
E4418B
Power sensor
cable
Signal Generator 8648A
Power splitter
11667B
Type N to SMA
adapter
Power sensor
8482A
SMA cable
SMA to BNC adapter
50W feed
through
Figure 56 Connecting Equipment for Maximum Frequency Check Test