Agilent Technologies 3000 X-Series Manual de servicio Pagina 64

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 156
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 63
3Capturing Data
64 Agilent 3000 Series Oscilloscopes User’s Guide
However, digital signals have frequency components above the fundamental
frequency (square waves are made up of sine waves at the fundamental
frequency and an infinite number of odd harmonics), and typically, for
1 Ghz bandwidths and below, oscilloscopes have a Gaussian frequency
response.
Figure 28 Theoretical Brick-Wall Frequency Response
f
S
f
N
-3dB
Attenuation
Frequency
0dB
Vista de pagina 63
1 2 ... 59 60 61 62 63 64 65 66 67 68 69 ... 155 156

Comentarios a estos manuales

Sin comentarios