3Capturing Data
86 Agilent 3000 Series Oscilloscopes User’s Guide
To store recorded waveforms
To store recorded waveforms:
1 Press Acquire
.
2 In the Acquire menu, select Sequence.
3 In the Sequence menu, press Mode to select “Storage”.
To select the start frame
1 In the Sequence menu (AcquireSequenceMode=Storage), select
Start Frame.
2 Turn the Entry
knob to select a number from l to 1000.
To select the end frame
1 In the Sequence menu (AcquireSequenceMode=Storage), select
End Frame.
2 Turn the Entry
knob to select a number from l to 1000.
To save a recording
1 In the Sequence menu (AcquireSequenceMode=Storage), select Save.
To load a recording
1 In the Sequence menu (AcquireSequenceMode=Storage), select Load.