Agilent Technologies 3000 X-Series Manual de servicio Pagina 143

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 156
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 142
Service 8
Agilent 3000 Series Oscilloscopes User’s Guide 143
23 Calculate the gain at the maximum frequency using the expression and
record it in the “Performance Test Record” on page 144:
For example, if (Vout @ Max Frequency) = 371 mV, (Vin @ Max
Frequency) = 427 mV and Gain @ 1 MHz = 0.993, then:
Record this value in the Calculated Gain @Max Freq column in the
Analog Bandwidth - Maximum Frequency Check section of the
“Performance Test Record” on page 144. To pass this test, this value
must be greater than -3.0 dB.
24 Move the power splitter from channel 1 to channel 2 and repeat steps 3
through 24 using channel 2 as the source.
Vin
max
3.65 mW 50× 427 mV==
Gain
max
20 log
10
Vout
max
()Vin
max
()
Gain
1MHz
----------------------------------------------------------
=
Gain
Max Freq
20 log
10
371 mV 427 mV
0.993
--------------------------------------------
-1.16 dB==