Displaying Data 2
Agilent 3000 Series Oscilloscopes User’s Guide 51
To use a digital filter
You can apply a digital filter to the sampled waveform data.
1 If the channel’s menu is not currently displayed, press the channel
button (1
, 2).
2 In the Channel menu, select Digital Filter.
3 In the Filter menu, press the Filter Type menu button to select between:
— Low Pass Filter.
— High Pass Filter.
— Band Pass Filter.
— Band Reject Filter.
4 Depending on the type of filter selected, select Upper Limit and/or
Lower Limit, and turn the Entry
knob to adjust the limit.
The horizontal scale control sets the maximum value for the upper and
lower limits.
To invert a waveform
You can invert a waveform with respect to the ground level.
1 If the channel’s menu is not currently displayed, press the channel
button (1
, 2).
2 In the Channel menu, select Invert to toggle between “ON” and “OFF”.
Figure 22 and Figure 23 show the changes before and after inversion.