8Service
144 Agilent 3000 Series Oscilloscopes User’s Guide
Performance Test Record
Table 21 DC Gain Test
Vertical
Sensitivity
Power
Supply
Setting
VDMM+ VDMM- VScope+ VScope- Calculated
DC Gain
Offset Gain
Test Limits
Channel 1
2mV/div ±6mV +0.96 to +1.04
5mV/div ±15mV +0.96 to +1.04
10 mV/div ±30 mV +0.97 to +1.03
20 mV/div ±60 mV +0.97 to +1.03
50 mV/div ±150 mV +0.97 to +1.03
100 mV/div ±300 mV +0.97 to +1.03
200 mV/div ±600 mV +0.97 to +1.03
500 mV/div ±1.5 V +0.97 to +1.03
1 V/div ±2.4 V +0.97 to +1.03
2 V/div ±6.0 V +0.97 to +1.03
5 V/div ±15.0 V +0.97 to +1.03