110 16800 Series Portable Logic Analyzers Service Guide
5 Troubleshooting
To verify pattern generator output
1 Connect one of the 10460-series data pods to the end of
the pattern generator Pod 1 cable.
2 From the Agilent Logic Analyzer application’s main
menu, select SetupMy Pattern Generator-1Bus/Signals....
3 In the Pattern Generator Setup dialog’s Buses/Signals tab,
assign channels to Pod 1.
Table 15 Equipment Required to Verify Pattern Output
Equipment Critical Specification Recommended
Model/Part
Oscilloscope >
500 MHz Bandwidth 54820A
Probe 500 MHz Bandwidth 1160A
Output Data Pod no substitute 10460A-series