Agilent Technologies 8133A Manual de servicio Pagina 143

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 188
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 142
Replacing Assemblies 6
16800 Series Portable Logic Analyzers Service Guide 143
To remove and replace the tray assembly
1 Perform previous procedures:
“To prepare the instrument for disassembly" on
page 116.
“To remove and replace the cover" on page 118.
“To remove and replace the front panel assembly" on
page 125.
2 Using a Torx T10 screwdriver, remove the 2 side screws
securing the tray assembly to the chassis.
3 Slide the tray assembly out far enough to disconnect the
hard drive and power supply cables. Use the release
tab(s) to help disconnect the power supply cables.
4 Slide the tray out of the chassis.
5 Reverse this procedure to replace the tray assembly.
T-10
screws (2)
Hard
drive
cables
Power supply
cables
Chassis
Vista de pagina 142