4- 22 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
:CALCulate:OFFSet:ACQuire
:CALCulate:OFFSet:ACQuire
Automatically sets the null offset value used for calculating the limit test data. The
null offset value will be the currently available value read by the
:CALCulate:DATA:LATest? or :SENSe:DATA:LATest? command. Or, it will be 0.0
if a currently available value does not exist.
Syntax :CALCulate[c]:OFFSet:ACQuire
Example :CALC:OFFS:ACQ
:CALC2:OFFS:ACQ
:CALCulate:OFFSet:STATe
Enables or disables the null offset function used for calculating the limit test data.
The null offset value is set by the :CALCulate:OFFSet or
:CALCulate:OFFSet:ACQuire command.
Syntax :CALCulate[c]:OFFSet:STATe mode
:CALCulate[c]:OFFSet:STATe?
Parameter mode 1|ON|0|OFF (default). Parameter data type is boolean.
mode=1 or ON enables the null offset.
mode=0 or OFF disables the null offset.
Query response mode <newline>
mode is 0 or 1, and indicates that the null offset is off or on, respectively. Response
data type is NR1.
Example :CALC:OFFS:STAT 1
:CALC2:OFFS:STAT?