B2900 Specific Error
Error 308 Emergency; High temperature4 status detected; Channel n
High temperature 4 status was detected in channel n. All channel output is changed
to 0 V and the all output switch is opened. Execute the *TST? command.
Error 311 Emergency; Abuse detected; Channel n
Abuse status was detected in channel n. All channel output is changed to 0 V and the
all output switch is opened. Execute the *TST? command.
Error 312 Emergency; F-COM(minus) abuse detected; Channel n
F-COM (minus) status was detected in channel n. All channel output is changed to 0
V and the all output switch is opened. Execute the *TST? command.
Error 313 Emergency; F-COM(plus) abuse detected; Channel n
F-COM (plus) abuse status was detected in channel n. All channel output is changed
to 0 V and the all output switch is opened. Execute the *TST? command.
Error 314 Emergency; Low sense(minus) abuse detected; Channel n
Low sense (minus) abuse status was detected in channel n. All channel output is
changed to 0 V and the all output switch is opened. Execute the *TST? command.
Error 315 Emergency; Low sense(plus) abuse detected; Channel n
Low sense (plus) abuse status was detected in channel n. All channel output is
changed to 0 V and the all output switch is opened. Execute the *TST? command.
Error 321 Emergency; SMU main power supply failure detected;
Channel n
SMU main power supply failure was detected in channel n. All channel output is
changed to 0 V and the all output switch is opened. Execute the *TST? command.
Error 322 Emergency; SMU positive power supply failure detected;
Channel n
SMU positive power supply failure was detected in channel n. All channel output is
changed to 0 V and the all output switch is opened. Execute the *TST? command.
Error 323 Emergency; SMU negative power supply failure detected;
Channel n
SMU negative power supply failure was detected in channel n. All channel output is
changed to 0 V and the all output switch is opened. Execute the *TST? command.