2- 30 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Command Summary
Controlling Source/Measure Trigger
To synchronize acquire actions (measurement)
Trigger source is set to the same mode.
Delay time is set to the same value.
Measurement wait time control is set to OFF.
Measurement ranging mode is set to the fixed mode.
Figure 2-7 Operation Example Using Trigger Delay and AINT Trigger Source
Example of AINT trigger source.
Trigger is generated after
completing all of the actions set by ͒
:TRIG<:ACQ|:TRAN|[:ALL]>:SOUR AINT.
:TRIG1:ACQ
:TRIG1:TRAN
SMU2
SMU1
time
time
V
V
:TRIG1:ACQ:DEL
:TRIG2:TRAN:DEL
Trigger Trigger
:TRIG2:ACQ
:TRIG2:TRAN
:TRIG2:ACQ:DEL
:TRIG1:ACQ:DEL
:TRIG2:TRAN:DEL
:TRIG2:ACQ:DEL
Measure
Measure
Measure
Measure