5- 12 Agilent B2900 SCPI Command Reference, Edition 2
Error Messages
B2900 Specific Error
B2900 Specific Error
Positive error numbers are Agilent B2900 specific errors, not standard SCPI errors.
Consult service for errors 111 to 140.
Error 111 Self-calibration failed; Voltage offset, item; channel n
Failed the voltage offset self-calibration specified by item and channel n.
Error 112 Self-calibration failed; Current offset, item; channel n
Failed the current offset self-calibration specified by item and channel n.
Error 113 Self-calibration failed; Voltage gain, item; channel n
Failed the voltage gain self-calibration specified by item and channel n.
Error 114 Self-calibration failed; Current gain, item; channel n
Failed the current gain self-calibration specified by item and channel n.
Error 115 Self-calibration failed; CMR DAC, item; channel n
Failed the CMR DAC self-calibration specified by item and channel n.
Error 121 Self-test failed; CPU communication, item; channel n
Failed the CPU communication test specified by item and channel n.
Error 122 Self-test failed; Fan status, item; channel n
Failed the fan status test specified by item and channel n.
Error 131 Self-test failed; SMU communication, item; channel n
Failed the SMU communication test specified by item and channel n.
Error 132 Self-test failed; CPLD access, item; channel n
Failed the CPLD access test specified by item and channel n.
Error 133 Self-test failed; Trigger count, item; channel n
Failed the trigger count test specified by item and channel n.