Vista de pagina 111
4 Sample Sessions: Analyzing an Existing Trace How to Add a Reflective Event
112 E6000C Mini-OTDR Users Guide, E0302
How to Set the Level-Markers for
Measuring Reflectance
Level Markers You now see 3 level-markers on the trace, and a message
box asking you to adjust the level-markers (Figure 38).
Below this message, you see 4 buttons labeled Zoom,
Marker, Cancel, and Ok.
Two level-markers are to the left of the Event, and are
joined by a regression line. A third level-marker is to the
right of the Event.
Figure 38 Level-markers for analyzing reflectance
3 Press S
ELECT to continue.
The Ok box above the trace is now highlighted.
You can now use your L
EFT and RIGHT cursors to
highlight any of the other buttons, and S
ELECT to select
the highlighted command.
How to Zoom while Adding an Event
If the level-markers are too close together, or outside the
screen area, you may want to change the horizontal and
vertical zooming. You do this as follows: