Test V. E6006A Power Meter Submodule D Single-Mode/Multimode Module Performance Tests
E6000C Mini-OTDR User’s Guide, E0302
261
Set up equipment 11 Set the DUT to 1310.00 nm.
12 Enable the laser source and the Agilent 8156A output,
and wait 3 minutes until the laser has settled.
13 Set the DUT to display power levels in W.
Note result in test record 14 Note the displayed measured value on the DUT in the
test record.
Test of the other wavelength
Repeat for other wavelength 15 Set the laser source to 1550 nm (nominal), and set the
attenuator and the DUT to 1550.00 nm.
16 Repeat steps 1 to 14 for this wavelength, replacing all
settings of 1310 nm/1310.00 nm by 1550 nm/
1550.00 nm.
NOTE The Reference Power Meter 81524A and the DUT are both
of the same type InGaAs. This means that the wavelength
dependencies are equal.
As long as both the Reference Power Meter and the DUT
are set to the same wavelength, the actual wavelength of
the source does not noticeably add to measurement
uncertainties, if the source is within a
±20 nm limit of the
measuring wavelength.
Total Uncertainty/Accuracy Test
NOTE Make sure that all equipment has warmed up, and all
connectors are clean.
Make sure that all patchcords are fixed to the table and
will not move during measurements.
Repeat each of the following steps for each of the
specified wavelengths.