D Single-Mode/Multimode Module Performance TestsOptional Test VI: E6007A Visual Fault Finder
268 E6000C Mini-OTDR Users Guide, E0302
Set up DUT 5 On the DUT, select Visual Fault Finder:
and allow to settle.
Note result in test record 6 Note the displayed power level on the Multimeter in the
test report.
Optional test: Center Wavelength
NOTE The laser has been vendor tested, and specifications are
purely typical. Therefore, this test is not mandatory.
Connect equipment 1 Connect the E6007A output to the Optical Spectrum
Analyzer using an 81101A patchcord, and two 81000AI
interface adapters:
ensure that the OSA is switched on and has warmed
up.
ensure that the E6000C is switched on and has
warmed up.
enable the E6007A (DUT).
Set up OSA 2 On the Optical Spectrum Analyzer:
press I
NSTRPRESET
press AUTO/MEAS and wait until End of
Automeasure is displayed.
choose U
SER and then select the type of source to be
measured as FP (for Fabry-Perot Laser).
To show the display in linear mode:
press M
ENU
select AMPT on the left side of the display
Select Mod CW
Select ON