Agilent Technologies 6002A Guía de usuario Pagina 257

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 320
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 256
Performance Test Form Sheets D Single-Mode/Multimode Module Performance Tests
E6000C Mini-OTDR Users Guide, E0302
257
Performance Test for the Agilent E6000C with Single-mode Modules Page 6 of 6
Model Agt. __________ Module Report No. __________ Date ___________
No. Test Description
III. Attenuation Deadzone 1300 nm Wavelength
Maximum Specification
Attenuation
Deadzone
Meas.
Uncertainty
Return Loss
35dB
Conditions:
Meas. Span. 2km
Pulsewidth 10ns
m
_______
m
10
m
_______
IV. Distance Accuracy 1300 nm Wavelength (Optional test)
Fiber Length: ________ m
Start Position: ________ m
Sample Spacing: ________ m (as
on the screen)
|Distance Accuracy| = (Fiber Length x Scale Error + Offset Error + 1/2 Sample Spacing)
|Distance Accuracy| = (_______ m x 10
-4
+1m + ________ m)
|Distance Accuracy| = ___________ m
Minimum Distance = Fiber Length - | Distance Accuracy|
Maximum Distance = Fiber Length +| Distance Accuracy|
Meas.Span
2 km
Pulsewidth
1
µs
100 ns
Minimum
Distance (typ.)
m
_______
Measured
Distance
m
_______
_______
Maximum
Distance (typ.)
m
_______
Meas.
Uncertainty
m
_______
Vista de pagina 256
1 2 ... 252 253 254 255 256 257 258 259 260 261 262 ... 319 320

Comentarios a estos manuales

Sin comentarios