The Settings screen 1 Getting Started
E6000C Mini-OTDR Users Guide, E0302
51
NOTE Changes made to the settings screen only affect
subsequent measurement acquisitions. However, you can
apply the changes that you have made to a trace that is
currently running by pressing R
UN/STOP again.
The Measurement Settings screen
The Measurement Settings page contains a list of
parameters that you can set (Figure 12).
Figure 12 The Measurement Parameters Settings page
You can change the following parameters:
Range Range: the start position and the measurement span. If
the Range displays Auto, the OTDR selects a suitable
measurement range for your fiber.
You can choose from one of the predefined ranges, or
select Range Input.. and input a range of your
choice.
Pulse Width PulseWidth: the length of the pulses launched by the
OTDR into the fiber. Short pulses improve resolution,
but longer pulses are required for higher dynamic on
long fibers.