Agilent Technologies N2783B Manual de usuario Pagina 20

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 45
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 19
20
Notice how closely output matches input. Graph
shows V
in
and V
out
when driven from a 25Ω
source.
The flat response means the waveform on the
scope screen will match the waveform at the
probe tip—across an entire 4GHz bandwidth.
Graph shows response (V
out
/V
in
).
Single-ended Active Probes - 1156A/57A/58A Active Probes
Operating characteristics
1156A/57A/58A
Bandwidth (–3 dB) 1156A: > 1.5 GHz; 1157A: > 2.5 GHz; 1158A: > 4 GHz
System bandwidth (–3 dB) 1158A with DSO/DSA90404A, DSO/MSO9404A: 4 GHz
1157A with DSO/DSA90254A, DSO/MSO9254A: 2.5 GHz
1156A with DSO/MSO9104A: 1 GHz
1156A with DSO/MSO9064A: 600 MHz
Rise and fall time (10% to 90%)
calculated from t
r
= 0.35/band-
width
1156A: < 233 ps; 1157A: < 140 ps; 1158A: < 88 ps
Input capacitance 0.8 pF
Input resistance* 100 kΩ 1%
Flatness, swept response 0.2 dB: 100 kHz to 100 MHz;
0.4 dB: 100 MHz to 2.5 GHz;
2.0 dB: 2.5 GHz to 4.0 GHz
Flatness, step response ±6.75 V dc to 100 Hz; ±1.25 V >100 Hz
10% overshoot: 75 ps input edge; 2%: 1 ns after edge
Dynamic range** > 5.0 V peak-to-peak
DC attenuation* 10:1 ± 3% before calibration****;
10:1 ± 1% after calibration****
Zero offset error referred to input* < 30 mV before calibration****;
< 5 mV after calibration****
Offset range* ±15.0 V
Offset accuracy* < 3% of setting before calibration****;
1% of setting after calibration****
Noise referred to input 3.0 mVrms
Propagation delay 5.5 ns
Maximum input voltage 40 V peak, CAT I***
ESD tolerance > 5 kV from 100 pF, 300 Ω HBM
Temperature drift Offset: < 1.0 mV/°C; Attenuation (Gain): 0.1 %/°C
* Denotes warranted specifications, all others are typical.
** For waveforms with edges > 3 ns, the dynamic range is > 12.0 V peak-to-peak.
*** Installation category (over voltage category) I: Signal level, special equipment, or parts of equipment,
telecommun nication, electronic, etc., with smaller transient overvoltage than installation category (overvoltage
category) II.
**** Probe calibrated to scope channel (under Probes Setup menu).
Vista de pagina 19