High Voltage Passive Probes - N2771B 1000:1 Passive Probe
Ideal for measuring up to 30 kV
Up to 250 MHz bandwidth
100:1 or 1000:1 attenuation
The N2771B is a 1000:1 divider probe for
the measurement of fast high voltage sig-
nals up to 30 kV dc + peak ac, 10 kV rms.
The probe’s large size and rugged
construction provide superior protection.
The ground lead is fed through the body
of the probe and protrudes behind the
safety barrier, keeping the ground connec-
tion away from the high voltage. Typical
applications include PMTs, motor drives,
high voltage switches, magnatrons, and
modern projection systems.
The N2771B is compatible with any
oscilloscope with 1 MΩ BNC input. For
use with the Infiniium 90000 Series scope,
use the E2697A high impedance probe
adapter. For use with the Infiniium 90000X
Series scope, use the N5449A high
impedance probe adapter.
N2771B high-voltage probe
N2771B derating curve
Operating characteristics the N2771B 1000:1 passive probe
Model
Number Description Quantity
N2771B High-voltage probe: includes alligator ground lead,
1-sharp-probe tip
1
Ordering information
N2771B
Bandwidth 50 MHz (-3 dB)
Risetime < 7 ns
Attenuation ratio 1000:1
Input resistance 100 MΩ (when terminated into 1 MΩ)
Input capacitance 1 pF
Compensation range 6 to 20 pF
Max. voltage 15 kV dc, 10 kV rms,
30 kV dc + peak ac
Operating temperature 0 to +50°C,
80% RH
Storage temperature -20 to +70°C,
90% RH
Dimensions 2 cm (max width of probe stem after handle) x 33 cm
7.5 cm (max probe widthat probe handle) x 33 cm
Probe readout No
Cable length 2 m
Vista de pagina 35
1 2 ... 31 32 33 34 35 36 37 38 39 40 41 ... 44 45

Comentarios a estos manuales

Sin comentarios