34
The Agilent 54006A allows you to probe sig-
nals up to 6 GHz using replaceable tips that
provide either 10:1 division ratio with 500Ω
input resistance, or a 20:1 division ratio with
1 kΩ input resistance. This 6 GHz probe
gives access to circuit nodes that are not
50Ω or do not have 50Ω connectors allow-
ing you to see signals at specific points, such
as the input to a gate. Agilent 54006A’s
input capacitive loading is approximately
0.25 pF, allowing you to get very accurate
timing measurements for a wide bandwidths
of signals.
The 54006A probe is a good, low-cost
alternative for high-frequency probing
where the higher resistive loading is not
an issue and the other features of the
InfiniiMax probing system are not needed
(such as differential inputs and multiple
connectivity options).
54006A for probing high frequency, up to 100 Ω impedance signals.
• Useful in probing high-frequency
signals with low source impedance
• Supplied with 10:1, 500 Ω and 20:1,
1 kΩ resistor dividers
• Low capacitive loading to
extremely high frequencies
General Purpose Passive Probes - 54006A High Bandwidth Passive Probe
54006A
Bandwidth (–3 dB) 6 GHz
Attenuation ratio 10:1, 20:1
Input resistance 500 Ω, 1 kΩ
Input capacitance 0.25 pF
Max dc volts 20 V
Length in meters (feet) 0.9 m (3 ft)
Operating characteristics the 54006A passive probe
Model
Number Description Quantity
54006A* 6 GHz Resistor Divider Probe
Includes:
One 10:1 500 Ω probe body, six 450 Ω resistors,
One 20:1, 1 kΩ probe body, six 950 Ω resistors,
One 36 in, 50 Ω coaxial cable, SMA (m-m)
One blocking cap, 10 GHz-26 GHz APC – 3,5 (m-f)
1
* Requires the 54855-67604 SMA to precision-BNC adapter to connect to BNC scope input.
Ordering information
Comentarios a estos manuales