AC/DC Current Probes - N2780B/81B/82B/83B AC/DC Current Probes
N2780B Series
Bandwidth (-3 dB) dc to 2 MHz (N2780B)
dc to 10 MHz (N2781B)
dc to 50 MHz (N2782B)
dc to 100 MHz (N2783B)
Maximum current (continuous) 500 A (N2780B)
150 A (N2781B)
30 A (N2782B/N2783B)
Maximum peak current (non-continuous) 700 A peak (N2780B)
300 A peak (N2781B)
50 A peak (N2782B/N2783B)
Maximum input voltage
300 V CAT I (N2782B, 83B)
300 V CAT III, 600 V
CAT II (N2780B, 81B)
Output voltage rate 0.01 V/A (N2780B/N2781B)
0.1 V/A (N2782B/N2783B)
Amplitude accuracy ±1.0 % rdg ±500 mA (N2780B)
±1.0 % rdg ±100 mA (N2781B)
±1.0 % rdg ±10 mA (N2782B)
±1.0 % rdg ±10 mA (N2783B)
Operating characteristics of the N2780B Series current probes
Model Number Description
N2780B 2-MHz current probe
N2781B 10-MHz current probe
N2782B 50-MHz current probe
N2783B 100-MHz current probe
N2779A Power Supply for the
N2780B/81B/82B/83B
Current Probes
Ordering information
The N2780B Series current probes are high
bandwidth, active current probes, featuring
flat bandwidth, low noise (2.5 mArms) and
low circuit insertion loss. In conjunction
with the power supply (model N2779A),
this probe can be used with any oscil-
loscope with a high-impedance BNC input.
The companion power supply N2779A
(3 x 12 Vdc output) lets you connect up to
any three N2780B-83B current probes to a
single power supply.
N2780B Series current probes with N2779A power supply
Vista de pagina 28