Agilent Technologies E6640A EXM Manual de servicio Pagina 142

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 340
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 141
142 Chapter 10
Programming a Sequence
Sequence Programming Example
Programming a Sequence
Use one command for first acquisition 1 setup
You can use one command to set up all the Acquisition parameters as follows:
:SENSe:LSEQuencer:ACQuire1:SETup
GSM,PGSM,MS,1,1,0,LOW,0.2ms,15.99ms,VIDeo,-10,0,NONE,-20
which sets up Acquisition 1 with the following
Radio standard: GSM
Radio band: PGSM
Device type: mobile station
Channel number: 1
Number of averages: 1
Peak power: dB
Instrument gain type: LOW
Transition time: 0.2 ms
Acquisition duration: 15.99 ms
Input trigger type: Video
Input trigger level: -10 dB
Input trigger delay: 0 ms
Output trigger: NONE
Instrument gain “LOW” value: -20 dB
Use one command for first analysis interval setup
You can use one command to set up the analysis interval parameters as follows:
:SENSe:LSEQuencer:ACQuire1:ASTep1:SETup 0ms,0.533ms,19,-5
which sets up Analysis Interval 1 with the following
Analysis step offset: 0 ms
Analysis step interval: 533 ms
Measurement bitmap: 19 (equivalent to binary 10011), which gives results for the following
measurements:
Basic Transmit Power
Basic Frequency and Phase
EDGE Power vs. Time
Expected power at the DUT output: -5 dB
Vista de pagina 141
1 2 ... 137 138 139 140 141 142 143 144 145 146 147 ... 339 340

Comentarios a estos manuales

Sin comentarios