98 Chapter 7
List Sequencer Spreadsheet Data Entry
Loading a tsv file into the List Sequencer
List Sequencer Spreadsheet Data Entry
Loading a tsv file into the List Sequencer
Once the tsv file is in the My Documents\Sequences directory on the test set it is imported into the List
Sequencer using the Recall hardkey menu.
Step 1. Press
Recall, Sequences
Step 2. Select from Source Sequence, Analyzer Sequence or Source and Analyzer Sequence.
Step 3. Select
Open....
Step 4. In the Open dialog box, select a sequence file and select Open (see Figure 7-3.). The selected
file is now ready for use in the list sequencer.
Figure 7-3. Load tsv file
Tab separated value file editing
The tab separated value (tsv) file which is used by the test set to set up the sequence can be edited using
a standard text editor. The tsv file is located in the test set on the D-drive at My Documents\Sequences.
The default name of the file is Sequence_xxxx.txt, where xxxx is a number that is automatically
appended when the file is saved. It is easier to edit the tsv file if it is imported into a spreadsheet. To do
this see “Importing a tsv file into a spreadsheet” on page 89.