Agilent Technologies E6640A EXM Manual de servicio Pagina 154

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 340
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 153
154 Chapter 11
Example Test Scenarios
Calibration Example
Example Test Scenarios
When the sequence has been completely set up, Sequence Studio can then save it in any of three formats:
a text file containing the sequence parameters as tab-separated values, a set of SCPI commands, or a
sequence file (.seq) in a proprietary format which is readable by the EXM test set. The first two formats
are illustrated below. (The SCPI commands for a sequence would typically be incorporated into a larger
test program which also controls the DUT.)
Figure 11-6. Sequence Exported as Tab Separated Values or SCPI Commands
In this example, the DUT-control application runs the sequence, collects the measurement data, and
saves calibration factors to the DUT to correct for any differences between expected and measured
power.
Vista de pagina 153
1 2 ... 149 150 151 152 153 154 155 156 157 158 159 ... 339 340

Comentarios a estos manuales

Sin comentarios