102 Chapter 4
Troubleshooting
Troubleshooting Using Diagnostic Test
Figure 4-10 Open dialog box
Step 4. Select "User [D]" (preset state) from menu in the "Look in:" box.
Step 5. Double-click "Agilent" folder to open it and to access its menu.
Step 6. Double-click "Service" folder to open it.
Step 7. Click "DiagnosticTest. VBA" program file to select it from program menu.
Step 8. Click "Open" button to download the diagnostic test program.
Step 9. Press
Select Macro to select Select Macro function.
Step 10. Press
Module1 main to open the Module1 main program file. "Diagnostic Test Program"
dialog box (Main Menu) will appear as shown in Figure 4-11.
Figure 4-11 Diagnostic Test dialog box
Step 11. To exit the diagnostic test, click "Exit" button.
Program Overall
Instrument
Information The option and serial number for the E5061A/E5062A, and test date
can be entered automatically.
Test Conditions The test conditions, “Temperature”, “Humidity” and person doing the
test (“Tested by”) can be entered.