52 Chapter 2
Performance Tests
E5061A Performance Test Record (for Option 175/275)
Crosstalk Test
Direction: S21
Direction: S12 (Option 275 only)
System Dynamic Range Test
For Serial Prefix : JP1KL or MY441
Direction: S21
Direction: S12 (Option 275 only)
For Serial Prefix : MY442 and above
Direction: S21
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
300 k - 100 M 10 -110
100 M - 1.5 G 10 -110
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
300 k - 100 M 10 -110
100 M - 1.5 G 10 -110
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
300 k - 100 M 10 -115
100 M - 1.5 G 10 -115
300 k - 100 M 3 k -90
100 M - 1.5 G 3 k -90
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
300 k - 100 M 10 -115
100 M - 1.5 G 10 -115
300 k - 100 M 3 k -90
100 M - 1.5 G 3 k -90
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
300 k - 1 M 10 -115
1 M - 100 M 10 -120