Installing your System
Flow Connections at the Isocratic Pump
Flow Connections at the Isocratic Pump
WARNING When opening capillary or tube fittings, solvents may leak out. Please
observe appropriate safety procedures (for example, goggles, safety
gloves and protective clothing) as described in the material handling
and safety data sheet supplied by the solvent vendor, especially when
toxic or hazardous solvents are used.
1 Remove the front cover by pressing the snap fasteners on both sides.
2 Place the solvent cabinet on top of the isocratic pump (Figure 10 on page 27.
3 Place the bottle containing your solvent into the solvent cabinet and place the
bottle-head assembly into the bottle.
4 Connect the solvent tube from the bottle-head assembly to the inlet adapter of
the active inlet valve. Fix the tube in the clips of solvent cabinet and isocratic
pump.
5 Using a piece of sanding paper connect the waste tubing to the purge valve
and place it into your waste system.
6 Connect the outlet capillary (isocratic pump to manual injection valve) to the
outlet of the purge valve.
Preparations Pump is installed in the HPLC system.
Parts required Other modules
Parts from accessory kit G1310-68705
Two wrenches 1/45/16 inch for capillary connections