Setting up an Analysis
Preparing the HPLC System
12 Click the button and select Change... to open the Signal Plot information.
Select the Isocratic Pump: Pressure and the VWD A: Absorbance at 254 nm
as signals. Change the Y-range for the VWD to 1 mAU. The X-axis range should
be 20 minutes. Press OK to exit this screen.
Figure 26 Edit Signal Plot Window
The Online Plot (Figure 27) shows both, the pump pressure and the VWD
absorbance signals. Pressing Adjust the signals can be reset to the offset
value and Balance would do a balance on the VWD.
Figure 27 Online Plot Window
Pump pressure signal
VWD absorbance signal