Installing your System
Installing the LAN Communication
Installing the LAN Communication
Your instrument HPLC system is operated via a standard LAN crossover
connection between the Value Solution ChemStation PC and the 1100
detector (VWD). You must ensure proper communication between the PC
and the analytical instruments. The communication uses the TCP/IP
protocol. For the configuration of the JetDirect card, that is used to connect
the analytical instrument to the LAN, the Boot Strap protocol is used, which
requires a Bootp Server.
NOTE The TCP/IP protocol and the Bootp Server program are already factory
pre-installed.
Configuring the CAG Bootp Server Program
NOTE Assure that the VW detector (with JetDirect card installed and connected to
the PC) is powered off.
1 The CAG Bootp Server program is placed in the start-up group and
automatically is started during the boot process of the PC. Its minimized and
located in the task bar.
2 Open the Bootp Server window by clicking on it in the task bar.
3 Now turn on the VW detector on the LAN and view the Bootp Server window.
After 2 - 10 seconds the Bootp Server will display the request from the
JetDirect card with the hardware (MAC) address (this information is also
stored in the file trace.txt in the bootp server directory, if Log to Disk is
enabled), see Figure 13.
The MAC or hardware address of the JetDirect card is a world wide unique
identifier. No other network device will have the same hardware address.