Setting up an Analysis
Before Using the System
Before Using the System
Solvent Information
Observe recommendations on the use of solvents in chapter Solvents in the
pumps reference manual.
Priming and Purging the System
When the solvents have been exchanged or the pumping system has been
turned off for a certain time (for example, overnight) oxygen will re-diffuse
into the solvent channel between the solvent reservoir, vacuum degasser
(when available in the system) and the pump. Solvents containing volatile
ingredients will slightly lose these. Therefore priming of the pumping system
is required before starting an application.
Table 5 Choice of Priming Solvents for Different Purposes
Activity Solvent Comments
After an installation
When switching between reverse phase
and normal phase (both times)
Isopropanol
Isopropanol
Best solvent to flush air out of the system
Best solvent to flush air out of the system
After an installation Ethanol or Methanol Alternative to Isopropanol (second choice)
if no Isopropanol is available
To clean the system when using buffers
After a solvent change
Bidistilled water
Bidistilled water
Best solvent to re-dissolve buffer crystals
Best solvent to re-dissolve buffer crystals
After the installation of normal phase
seals (P/N 0905-1420)
Hexane + 5% Isopropanol Good wetting properties