53
Troubleshooting your System
Diagnostic Functions
Diagnostic Functions
Status Information on the 1100 Modules
LEDs
The 1100 modules provide two LEDs each.
• A green LED is built into the power switch and indicates that the power
supply has been turned on. It does not indicate, whether all voltages are
present or not.
• A multi-color LED at the top-right of the module, indicates the actual state
of the module/system.
RED: indicates an error condition in the module/system (e.g. leak).
YELLOW: indicates a not-ready condition (e.g. lamp is off)
LED OFF: the system is ready (ChemStation is GREEN)
GREEN: the system is in run mode (ChemStation is BLUE)
In case of an error condition, check for corresponding error messages in
the system log or instrument error log, see Figure 31 on page 56.
Analog Signals
The isocratic pump and the variable wavelength detector provide an analog
output, where module specific analog signals (pump: pressure, pressure
ripple; VWD: absorbance signals, board temperature, …) can be selected and
plotted to a connected recording device or routed into a data system (via A/D
converter).
NOTE For detailed information on error messages, signals and tests refer to the user
interface help system and the module’s Reference Manual, section
Diagnostics.