Triggering Examples
Special Displays
To interleave trace lists
1 Set up both analyzers as state analyzers.
2 Go to the Trigger menu of the first analyzer.
3 Set Count to Time, and set up the trigger.
The logic analyzer uses the time tags stored with each state to
determine the ordering of states shown in an interleaved trace list.
4 Set Count to Time, and set up the trigger on the second
analyzer.
The second analyzer does not need to be run from the first analyzer,
but both analyzers must have "Count Time" turned on to correlate the
data.
5 Make the measurement run.
6 Go to one of the Listing menus.
7 Select one of the label fields in the trace list display, then select
Interleave.