Retrieving and Restoring Data Using the LAN
Retrieving and Restoring Data Using the LAN
To restore configurations
1 Access the logic analyzer's file system as the control user.
Refer to the chapter "Accessing the Logic Analyzer File System".
2 Copy the setup.raw file to the appropriate directory.
For system configurations, this would be the \system directory. For
analyzer configurations, this would be the \slot_a directory. For
oscilloscope configurations, it would be the \slot_b directory. If you
copy a configuration file to a different directory than it came from, the
analyzer will not know how to interpret it.
Example Restoring an analyzer configuration
Suppose you want to load a configuration file called "486_bus" from
your local computer into the logic analyzer's analyzer. The analyzer is
always in slot A. The logic analyzer is mounted on your network as disk
drive L:.
To load the configuration file, at the MS-DOS prompt enter:
copy 486_bus L:\slot_a\setup.raw
If your computer is running the UNIX operating system, you might use
the cp command. In an MSWindows environment, you can use File
Manager.